Soxtherm Glass beakers
Recirculating coolers
Recirculating coolers

Recirculating cooler

FL

The recirculating cooler is especially designed to work with Gerhardt analysis systems, like the VAPODEST product line, the SOXTHERM models as well as the digestors SMA (trace metal) and CSB (chemical oxygen demand).

This cooler may be used for continuous operation in combination with the Gerhardt systems as it offers constant working temperature as well as consistent pressure without contamination of the cooling agent.The multi-display ensures a user-friendly and easy operation. Optimal safety features are the automatic self-test when starting the instrument combined with sensors for the operation as well as protection of the pumps and the compressor. Furthermore, a high and low temperature warning function as well as a freezing protection are part of the features of the recirculating cooler. The space saving design allows installation underneath the lab bench.

Models available and possible combination

Recirculating cooler FL 1201

designed for continuous operation in combination with

Recirculating cooler FL 601

designed for continuous operation in combination with


Technical specifications FL 601 and FL 1201

Nominal voltage 230 VAC 50/60 Hz
Working temperature range - 20 to + 40 °C
Temperature stability +/- 0.5 °C
Display / Resolution LED / 0.1

Pump capacity max.
Pump pressure

40 l/min
1000 mbar
Cooling capacity (20 °C)

600 W (FL 601) resp.
1200 W (FL 1201)

Ambient temperature allowed 5 to 40 °C
Cooling agent R 404 a
Weight

45 kg (FL 601) resp.
69 kg (FL 1201)

Dimensions (W x D x H) 320 x 500 x 600 mm (FL 601) resp.
500 x 760 x 640 mm (FL 1201)
 © 2008-2011 C. Gerhardt GmbH & Co. KG
Gerhardt Analytica Systems